testo 176 H1

  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
  • testo 176 H1
TESTO AG
4-channel temperature and humidity data logger with external sensor connections (NTC/capacitive humidity sensor)
 
Part #: 0572 1765

Description

When short-term measurements show no anomalies, but the storeroom climate still does not meet the desired requirements, then the testo 176 H1 is the right data logger. It has two connections for external humidity probes which can be positioned in the room according to the individual requirements.
 
  • Large memory for 2 million measurement values
  • up to 8 years battery life
  • Standard battery (AA) replaceable by user
  • large, easily legible display
  • Humidity / Temperature Probe 12mm, L=1.3 m

    # 0572 6172
    Meas. range -20 to +70 °C, 0 to +100 %RH
    Accuracy   ±0,3 °C / ±2 %RH at +25 °C (2 to 98 %RH) / ±0,03 %RH/K
  • Humidity / Temperature Probe 4 mm, L= 2,2 m

    # 0572 6174
    Meas. range 0 to +40 °C, 0 to +100 %RH
    Accuracy ±0,3 °C / ±2 %RH at +25 °C (2 to 98 %RH)    ±0,08 %RH/K ± 1 digit
  • USB connection cable, instrument to PC

    # 0449 0047
  • ComSoft Professional 4

    # 0554 1704

    (Pro software incl. data archiving)

  • ComSoft CFR 21 Part 11

    # 0554 1705

    (Software for requirements according to CFR 21 Part 11 for Testo data loggers)

  • Mobile printer for data loggers testo 175/176/184

    # 0572 0576

    Thermal printer with USB/NFC interface. USB connection between testo 175 / testo 176 and the printer. NFC connection between testo 184 and the printer 

  • ComSoft Basic 5 CD software

    # 0572 0580

    (Basic software for programming and readout of Testo data loggers; graphic and tabular measurement data presentation as well as export function.)

  • ISO cal. cert./humidity

    # 0520 0076

    (humidity data logger; calibration points 11.3%RH and 75.3%RH at +25°C)

  • DAKKS calibration cert./humidity

    # 0520 0246

    (humidity data logger; cal. points 11.3%RH and 75.3%RH at +25°C)

Memory

2.000.000 readings

Storage temperature

-40 to +85 °C
Operating temperature                                        -20 to +70 °C

Dimensions

103 x 63 x 33 mm
 
Protection class                                                    IP65
Measurement rate                                               1 sec to 24 h
Battery life                                                           8 years at 15 min. measurement rate
 
 
Probe type Typ T (Cu-CuNi)
 
Meas. range                                                        -200 to +400 °C
Accuracy ±1 %of mv (-200 to -100,1 °C)
±0,3 °C (-100 to +70 °C)
±0,5 %of mv (+70,1 to +400 °C)
Resolution 0,1 °C
 
 
Probe type NTC
 
Meas. range -20 to +70 °C
Accuracy ±0,2 °C (-20 to +70 °C)
±0,4 °C (remaining range)
Resolution 0,1 °C
 
 
Probe type Type K
 
Meas. Range -195 to +1000 °C
Accuracy ±1 %of mv (-200 to -100,1 °C)
±0,3 °C (-100 to +70 °C)
±0,5 %of mv (+70,1 to +1000 °C)
Resolution 0,1 °C
 
 
Probe type humid. sensor, cap.
 
Meas. Range 0 to 100 %RH
Resolution 0,1 %RH
 
 
Probe type Typ J (FeCuNi)
 
Meas. Range -100 to +750 °C
Accuracy ±0,3 °C (-100 to +70 °C)
±0,5 %of mv (+70,1 to +1000 °C)
Resolution 0,1 °C
 